Spectronix Ar 2015
The device is capable of providing fast and simultaneous spectral collection in UV-Vis region.




   The benefits of array technology include very fast scanning
   with ability to scan all wavelengths range of 200-800 nm
   (190-1100 nm) in less than 1second,
Fiber optic connectivity
   is one of the features of this device, which leads to the           

   flexibility of the device and increased its use.   
 







   Optical Specifications     Description
   Wavelength range     200-800 nm (190-1100 nm)
   Straylight    < 0.03 %
   Resolution    < +1.0 nm
   Wavelength accuracy    < ± 0.1 nm
   Wavelength reproducibility    < ±0.02 nm for 10 consecutive scans (NIST 2034)
   Photometric accuracy    < ±0.01 A at 440.0 465.0 590.0 635.0 nm (NIST 930 e)
   Photometric noise    < 0.01 A 80 scan at 0 A , 500 nm
   Photometric stability    < 0.01 A/min at 0 A , 500 nm
   Baseline flatness    < 0.01 A 0.5 Second blank ,0.5 second scan at 0 A , 500 nm
   Scan time    2-2000 ms
   Typical scan time    0.1 s
   Light source    Deuterium+Halogen / Tungsten lamp
   Detector    2000 Multichannel
    Physical Specifications     Description
   Communication    USB 2
   Software    Windows Xp ,7,8
   Height * Width * depth    22 * 12 * 27.5 cm
   Weight    4 kg
   Power    220 V AC 50/60 Hz

Applications:

This device is for analysis and the spectrum of sodium vapor lamps, Incandescent, LED, Neon, hot material optical spectrum, spectrum analysis of color filters. These types of spectrometers are used in chemical analysis and particle physics.






Research Lab
Teaching Lab
Chemistry Life Science

Views count : 534